What does ISO 14971 require for risk management in medical software?
Short answer
An Ongoing Process Rather Than a One-Time FMEA
ISO 14971:2019 describes risk management from the initial product idea to decommissioning. For medical software, this means that risks are not only collected shortly before CE marking but are linked with requirements, architecture, testing, changes, cybersecurity, and market observation. Annex I number 3 of the MDR also requires a continuous, iterative risk management system for each product.
At the beginning is a risk management plan. It defines responsibilities, review steps, criteria for risk acceptance, methods for assessment, as well as the collection and evaluation of production and post-market data. The criteria must be defined before the assessment; the standard deliberately does not specify a universal acceptable numerical value.
From Hazard to Residual Risk
The core process according to ISO 14971:2019 includes:
- describe intended use and reasonably foreseeable misuse,
- identify safety-related characteristics and hazards,
- model hazardous situations and potential harms,
- assess and evaluate risks based on established criteria,
- select, implement, and verify the effectiveness of risk control measures,
- evaluate remaining individual risks and the overall residual risk,
- weigh benefits against residual risk if necessary,
- continuously evaluate production and post-market information.
The control hierarchy starts with inherently safe design. Only after that come protective measures and finally safety information. In software, this can mean, for example: architecturally preventing hazardous states, validating inputs, creating a safe fallback level, and clearly communicating remaining limitations. A warning notice does not replace a technically feasible risk reduction.
Traceability Makes the File Robust
Every control should be traceable to the affected hazard, a requirement, and a verification proof. New insights from support, security vulnerabilities, error reports, or clinical use can change the original assessment and must flow back into the risk file, testing, and, if applicable, clinical evaluation and user information.
In the EU, EN ISO 14971:2019 including A11:2021 is listed as a harmonized standard. Its application is voluntary; it can provide a presumption of conformity for covered MDR requirements. It does not replace the MDR or a product-specific assessment but offers the recognized process framework.
Example from practice
In a dosing software, not only "incorrect calculation" is noted. The cause, output value, clinical usage context, potential harm, risk control, and the test that demonstrates the effectiveness of this control are documented.
Key facts
- Current Edition
- ISO 14971:2019, 3rd Edition
- EU Version
- EN ISO 14971:2019/A11:2021
- Duration of Validity
- Entire product lifecycle
- MDR Reference
- Annex I number 3 MDR
Sources
All external claims are backed by traceable sources.-
01
ISO 14971:2019 – Application of risk management to medical devices International Organization for Standardization (ISO)
-
02
Durchführungsbeschluss (EU) 2022/757 zu harmonisierten Normen EUR-Lex / Europäische Union
-
03
Verordnung (EU) 2017/745, insbesondere Anhang I Nummer 3 EUR-Lex / Europäische Union